Product Specifications:

ItemUniversal Laboratory Penetrometer
Complies WithASTM D5,D217,D1168,D1191,D1321, D1403,D1831,D1855, AASHTO T49,T187 and others
Product Datasheethumboldt-h-1200.pdf
RangeDependent on Penetration Needle or Cone Used

Shipping Info:

Ship weight (kg) 9.9945
Ship height (cm)55.37
Ship length (cm)38.1
Ship width (cm)36.83
HS code9027808060
Country of OriginUS

Product Details:

Humboldt H-1200 universal laboratory penetrometer accurately measures the penetration of cement, bituminous materials, petrolatum and waxes in food, cosmetics & pharmaceutical applications.

Features:

  • Humboldt H-1200 universal penetrometer is equipped with a 5-inch diameter dial indicator having 400 divisions of 0.1 mm for a maximum of 40 mm penetration in one revolution of the dial.
  • This dial indicator can be preset to zero to eliminate errors.
  • The universal penetrometer also comes with a 47.5 g plunger having a 3.2 mm hole and two loading weights of 50 g & 100 g.

Standards and Approvals:

  • ASTM D5
  • ASTM D1168
  • ASTM D1321
  • ASTM D937
  • ASTM D217
  • AASHTO T49, AASHTO T187

Frequently Asked Questions:

Q. How to install this universal penetrometer?

A.

  • Level the unit carefully using levelling screws present on the base and maintain proper alignment using a spirit level.
  • Attach a suitable penetrating needle / cone into the unit's chuck.
  • Set the zero reference. Use the zero setting screw if required. Release the clutch trigger by holding the gear rack knob & pulling it up. Raise the test plunger to its maximum height after the trigger is released.

Q. What are some safety points to be remembered while operating this universal penetrometer?

A. Users should keep their body parts, clothing and other necessary objects away from the moving parts when the instrument is in operation.